Example:
Send TCP/IP command to Model 9016 module (via its connected socket), that
returns decimal pressure data for channels 1, 5, 9, and 13 in ASCII fixed point
format:
r11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
1.234000 0.989500 1.005390 0.899602
This command example also works for Models 9021 and 9022 if the bits in the
position field are restricted to channels 9, 5, and 1. (“r01110”)