5.1.6.3 Adapter Plate O-Ring Replacement
Following is a step-by-step procedure to replace Adapter plate O-rings in a Model 9016
Intelligent Pressure Scanner. The adapter plate is located opposite of the tubing plate on the
calibration manifold. All DH200 transducers are attached to the adapter plate.
(1) Disassemble the module as described in Section 5.1.2.
(2) Remove the PC-206 Amplifier/Multiplexer board as described in Section 5.1.3.1. Lay
the circuit board to the side on an anti-static surface.
(3) Remove the six (6) 3/32" Allen-head screws that secure the adapter plate to the calibration
valve housing. To remove the two (2) center screws, you must remove the DH200
transducers near the screws. Make sure to note the DH200 serial number and location.
The plate should be gently lifted from the calibration housing.
(4) Carefully rotate or slide the adapter plate back and forth, pivoting on the guiding pin about
1/8" several times. This is done to loosen the O-rings from the calibration manifold. Lift
the adapter plate straight up. Do not touch the calibration manifold.
(5) Remove and replace the O-rings needing maintenance using the procedure described in
Section 5.1.6.1. Note that the O-ring seals use an additional PTFE cup seal placed on top
of the O-ring. These PTFE seals do not require Krytox
®
grease.
(6) Examine the adapter plate and calibration valve surface to verify that no contaminants are
on either surface. This generally requires microscopic examination. Replace the adapter
plate by slowly placing the plate on the calibration manifold. Make sure that the O-ring
is down towards the pneumatic sliding manifold and the guiding pin on the adapter plate
fits into the mating hole of the calibration valve housing. Fasten the adapter plate evenly
on all sides.
(7) Install the DH200 transducers that were previously removed. It is suggested to install
them back in their original location.
(8) Replace the PC-206 board as described in Section 5.1.3.1 and reassemble the module.
(9) Test your scanner to ensure proper operation.