Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
2 - 13
NOTE: The factory default setting for jumper JB1 is 10 VDC
Figure 2.5
9022 Jumper (JB1) Set For 10 VDC Excitation
2.3.8 Case Grounding
The 9016 and 9021/9022 modules contain a case bypass capacitor which allows the module case
to be mounted on hardware with a small common mode line voltage (less than 20 Volts).
2.3.9 Trigger Input Signal
Models 9016, 9021, and 9022 each support the use of a data acquisition synchronization signal,
sometimes called “Hardware Trigger.” When configured through the Define/Control Host
Stream (‘c’) command, the trigger signal can be used to initiate and synchronize data acquisition
and stream outputs to the host.
The trigger signal is intended to be a 2-wire differential signal brought in through the 9016, 9021,
or 9022 main electrical connector. The signal may be driven by a standard TTL compatible
device. The switching threshold for this signal is set at 2.5 VDC.
2.3.10 Power Up Checks and Self-Diagnostics
Upon power-up of the module, the internal firmware will perform a number of self-diagnostic
checks. The results of these tests are reflected by the ‘OK’ LED on the top panel. The 9016,
9021, and 9022 modules complete the power up and self diagnostics in approximately 30 seconds.
See Chapter 6, Troubleshooting Guide for additional information and potential problem areas
during the power-up sequence.