Pressure Systems 9022 Manual de usuario Pagina 89

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 194
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 88
Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 59
Example:
! Send TCP/IP command to Model 9016 module (via its connected socket), that
returns decimal pressure data for channels 1, 5, 9, and 13 in ASCII fixed point
format:
r11110
Response contains data for channels 13, 9, 5, and 1 (left to right):
1.234000 0.989500 1.005390 0.899602
This command example also works for Models 9021 and 9022 if the bits in the
position field are restricted to channels 9, 5, and 1. (“r01110”)
Vista de pagina 88
1 2 ... 84 85 86 87 88 89 90 91 92 93 94 ... 193 194

Comentarios a estos manuales

Sin comentarios